logo
CML HistoriesAdd CML History

Add CML History

Record a thickness measurement reading against a CML in SHIELD to build corrosion history and update remaining life calculations.

Overview

CML history records store individual thickness readings taken at a Corrosion Measurement Location over time. Each reading you add contributes to the CML's measurement history, which SHIELD uses to calculate corrosion rate and remaining life for the parent component. Accurate date, inspector, measurement method, and thickness data are all essential for reliable integrity analysis.

The inspector must already exist in SHIELD before they can be selected on a history record. If the inspector is not in the dropdown, add them first under Settings → Inspectors. See Add Inspectors for instructions.


Steps

Navigate to the Asset and Open CML Editing Mode

Go to the Assets module and open the asset you want to add a thickness reading for. Click the edit icon on the CMLs header to enter CML editing mode.

Navigate to the asset and open CML editing mode
Navigate to the asset and open CML editing mode

Click on the CML to Expand Its History

Locate and click the CML row you want to add a reading for. This expands the CML to reveal its measurement history list and the controls to add a new entry. Confirm the CML ID and description match the measurement location where the reading was taken.

Click the CML row to expand its history
Click the CML row to expand its history

Click the Add Icon for a New History Reading

Click the Add icon in the CML history section to open a new measurement entry row. Each click adds one reading entry — use this for each individual thickness measurement taken at this CML.

Click the Add icon for a new CML history reading
Click the Add icon for a new CML history reading

Enter the Inspection Date

Enter the date the thickness measurement was taken in the field. Use the actual inspection date — not today's date — to ensure corrosion rate calculations and trending are accurate over time.

Enter the inspection date
Enter the inspection date

Select the Inspector

Use the inspector dropdown to select the person who performed the measurement. Only inspectors already configured in SHIELD will appear in the list. If the required inspector is not listed, add them under Settings → Inspectors before continuing.

Select the inspector from the dropdown
Select the inspector from the dropdown

Enter the Measured Thickness

Enter the actual thickness reading obtained at the CML in the configured unit of measure (typically inches or mm). SHIELD will compare this value against previous readings for the same CML to calculate the corrosion rate and flag any significant wall loss or anomalous growth.

Enter the measured thickness reading
Enter the measured thickness reading

Complete Measurement Method and Device Information

Enter the remaining measurement details including method (e.g., UT straight beam, RT) and device information (e.g., instrument model or calibration reference). This data supports audit traceability and QA/QC verification of the reading.

Enter measurement method and device information
Enter measurement method and device information

Click Save

Click Save to record the thickness reading. SHIELD will update the CML's measurement history and recalculate the corrosion rate and remaining life for the parent component using the new data point.

Click Save to record the CML history reading
Click Save to record the CML history reading

Field Reference

FieldDescriptionRequired
Inspection DateDate the measurement was physically taken✅ Yes
InspectorPerson who performed the measurement✅ Yes
Measured ThicknessActual thickness reading in configured units✅ Yes
MethodInspection technique (e.g., UT, RT)✅ Yes
Device / InstrumentEquipment used — model or calibration reference✅ Yes

Troubleshooting

SymptomCauseFix
Inspector not in dropdownInspector not yet added to SHIELDAdd the inspector under Settings → Inspectors — see Add Inspectors
Corrosion rate not updating after saveOnly one reading exists — rate requires two data pointsAdd at least one prior baseline reading so SHIELD has two points to calculate rate from
Thickness reading flagged as anomalousLarge variance from previous readingVerify the reading is correct; if valid, add a note to the record explaining the deviation
CML history not visible after savePage not refreshedRefresh the asset record to confirm the reading has been added